Sold
This item has been SOLD.
Inventory No.
0321-024-0015
Manufacturer
Varian
Model No.
979
Dimensions L x W x H (inches)
Unavailable
Capacity
-
Electrical
1/50/60/115/230
Weight (lb.)
195
Category
Condition
See Details
Quantity
1
Serial No.
LLB9029
Make Year
-
Item Details
W/probe (bofo002), *contains 3 Pieces 0321-024-0015
About This Item
Product Overview:
Even with today's complex technology, it is, for all practical purposes, impossible to manufacture a sealed enclosure or system that can be guaranteed leak proof without first being tested. Through the use of modern mass spectrometer leak testing techniques, as implemented by the Varian 979 Helium Leak Detector, leak rates in the 10โ9 std cc/sec range can be reliably detected.
The Agilet 979 is a wide-range Helium Mass Spectrometer Leak Detector. It is comprised of a turbo-molecular high vacuum pump, spectrometer tube, valve block, Varian Vacuum Technologiesโ Platform leak detector electronics, and an operator interface in a stylish, yet rugged housing.
Product Features:
โข Agilent Technologies / Varian
โข Helium Leak Detector
โข Fully Re-manufactured
โข Roughing Pump not included
General:
Manufacturer: Varian
Model Name: Helium Leak Detector
Model Number/MPN: 979
Inventory Number: 0321-024-0015
Series: LLB9029
Power Requirements:
Power: 115V, 50 Hz | 115V, 60 Hz (20 amps)
Specifications:
Aisle: 03B
Bay: 3600
Weight (lb): 195 lbs.
Sensitivity Ranges in Atmospheric-Cc/sec Helium:
Std. Sensitivity, 1 Pri. Pump: 10-4 thru 10-9
Std. Sensitivity, 2 Pri. Pumps: 10-1 thru 10-9
High Sensitivity, 1 Pri. Pump: 10-5 thru 10-10
High Sensitivity, 2 Pri. Pumps: 10-2 thru 10-10
Minimum Detectable Leak at 1000 Ppm Ambient Helium:
Std. Sensitivity: 5 x 10-10 atm-cc/sec helium
High Sensitivity: 5 x 10-11 atm-cc/sec helium
Maximum Test-Port Pressure:
1 Primary Pump: 5 torr
2 Primary Pumps: 760 torr
Test-Port Connector: NW-25 KF flange (adapter to 1-1/8 in. compression fitting available)
Leak Rate Indication: 50-segment log or linear LED bar graph plus numeric readout on screen
Leak Rate Units Displayed: atm-cc/sec, mbar-l/sec, torr-l/sec, Pa-mยณ/sec
Test-Cycle Control: Automatic
Calibration Routine: Automatic or manual
Background Suppression (zeroing): Automatic and Auto-Zero < Zero
Changing Ranges: Automatic
Protection from Air-Inrush and Power Failure: Automatic
Audio Signal: Both threshold and volume can be controlled
Remote Analog Leak-Rate Output: Phone jack, 0 to 5V dc
Recommended Ambient Operating Temperature:
With Rotary-Vane Pump(s): 12 to 40 deg C
With Dry Scroll Pump(s): 5 to 40 deg C